The LIB benchtop environmental chamber is equipped with a new generation of intelligent PLC controllers, designed for precise environmental simulation testing:
1. Multi-language operation interface: Support English/Spanish/Russian/German and other multi-language switching, to meet the needs of global users.
2. Large capacity data storage: built-in 6GB storage space, can record 100,000 sets of test data (temperature/time/number of cycles).
3. Intelligent programming system: support create 120 programs 100 segments, each segment can be independently set temperature, humidity, dwell time and other parameters
LIB benchtop environmental chamber controller also includes real-time monitoring function and security protection mechanism.
A. Select "Program" mode:
Select “Program Settings” on the main interface to enter the Program Settings screen , and set the program and test name.
In the program setting screen, click "Program setting" to enter the program editing screen, and set the temperature, humidity, hours, minutes, and seconds required for the program group.In the program group editing, TS1, TS2, TS3 are waiting, TS4, TS5 are 0 by default , and no setting is required.
the program setting screen [1-Program setting screen] , click “Repeat Setting” to enter the loop setting screen, as shown in [3-Program setting screen], and set the loop count required for the program group.
☞Group loop: All programs in the program group are looped.
☞Loop count: Set the number of times the program group will be looped. Note that if you enter " 0 ", it will be repeated infinitely.
☞ Group connection: After the set program group is finished running, the program group that will run continuously. The default value is 0, no setting is required.
☞Segment loop: loop some segments in the program group. The default is 0, no setting is required.
B. Click "Monitoring Screen" on the main interface to enter the program mode monitoring screen. Click "Pattern No. " on the right to select the program group that has been set.
C. After setting the above test parameters, click the [Run] button in the lower right corner of monitoring screen to start the test.
It adopts compact design(options:50L,80L)for laboratory and small-batch testing:
Temperature control range | -20℃,-40℃,-70°C~+150°C (optional) |
Humidity range | 20%~98%RH |
Temperature uniformity system: | 3D circulating air duct design,temperature difference ≤±0.5°C |
Energy-saving design | R134a environmental refrigerant |
Compatible expansion | Ethernet interface for remote monitoring |