Xenon Test Chamber Features and Benefits
1. The use of xenon arc lamp as a light source, can simulate the full sunlight spectrum, including ultraviolet, visible light and infrared and other wavelengths of light.
2. Compared with the natural aging test, the xenon lamp test chamber can accelerate the aging process of the material and simulate the effect of long-term exposure of the material in the natural environment in a shorter time.
3. In addition to light, xenon test chamber can also accurately control temperature, humidity and water spray playing the role of weatherometer, the interaction between light and temperature, humidity and rain reproduce more authentic environment.
4. Color LCD touch screen controller, convenient for operators to set temperature, humidity, irradiance, spray, sample holder rotation and other parameters, to achieve automatic cycle test, improve test efficiency and repeatability.
Xenon Test Chamber Specifications
Model | XL-S-750 |
Internal Dimension (mm) | 950*950*850 mm |
Overall Dimension (mm) | 1400*1400*1950 mm |
Specimen Holder Size(mm) | 95*200 |
Specimen Capacity | 42pcs |
Chamber Type | Rotating Holder |
Irradiation Source | 1 piece of 4500w water-cooled xenon arc lamp |
Irradiance Range | 35~150 W/㎡ |
Bandwidth Measurement | 300nm~400nm (340nm or 420nm) |
Chamber Temperature Range | Ambient ~ 100 ℃ ±2℃ |
Black Panel Temperature | BPT 35 ~ 85 ℃ ±2℃ |
Humidity Range | 50% ~ 98% RH |
Humidity Deviation | ± 5% RH |
Water Spray Cycle | 1~9999H59M (Adjustable) |
Controller | Programmable color LCD touch screen controller |
Water supply system | Automatic water supply, Water purification system |
Radiometer | UV Radiometer, Tolerance: ±5% |
Cooling System | Mechanical compression refrigeration system |
Configuration of Xenon Test Chamber
① Workroom | Adopts SUS304 high quality stainless steel mirror panel, corrosion resistance, high temperature resistance, high reflectivity, can effectively reflect the light source of xenon lamp The rotating sample holder can make the test sample evenly exposed to light during the test,mounting 42 pieces of specimen. The water-cooled xenon lamp emit the full spectrum of light close to natural sunlight, and average lamp life is 1600 hours, 4500W/6500W. A variety of filtration systems are available, including glass Windows and UV filters.
Equipped with temperature and humidity sensors, irradiance sensor including full spectrum radiometer: 280~800nm and UV irradiance radiometer: 300~400nm,340nm,420nm. The sample holder is provided with 3 layers of nozzles to simulate rainfall or dew.
|
|
② Air circulation system | Centrifugal fan is used to make the air flow in the working room to ensure the uniform temperature and humidity in the workroom. | |
③ Heating system | Adopt nickel-chromium alloy electric heating heater. | |
④ Control system | PLC programmable controller, touch screen can set light intensity, temperature, humidity, rainfall cycle, monitor the test process and view the test results through the control system. | |
⑤ Humidity system | External isolated humidifier, all stainless steel shallow surface evaporation humidifier, can provide a stable humidity environment. | |
⑥ Refrigerating system | The refrigeration unit uses compressors for mechanical refrigeration. | |